Please use this identifier to cite or link to this item:
http://hdl.handle.net/2289/3213
Title: | Imidazole modified C-paste electrode in the determination of trace amounts of Mercury |
Authors: | Kannan, R. Rajagopalan, S.R. Lakshminarayanan, V. |
Issue Date: | 1995 |
Publisher: | The Electrochemical Society of India |
Citation: | Journal of the Electrochemical Society of India, 1995, Vol.44, p90-96 |
Abstract: | The imidazole-modified C-paste electrode was shown lo accumulate mercury (II) ions non- electrochemically from nitrate containing solutions at pH 7.0-7.3. The accumulated mercury gave well defined voltammograms, when subjected to voltammetric scale in a medium of potassium thiocyanate and perchloric acid; these are suitable for determination of mercury. In comparison with pure C-paste electrodes, the imidazole-CME is shown to offer improved resistance to interference by diverse types of ions present in natural waters. The method offered a sensitivity of 1.75 μA/ppm and a detection limit of 0.1 ppm (lor 15 min. accumulation time). The relative standard deviation was 12%. |
Description: | Restricted Access. |
URI: | http://hdl.handle.net/2289/3213 |
ISSN: | 0013-466X |
Copyright: | 1995 The Electrochemical Society of India |
Appears in Collections: | Research Papers (SCM) |
Files in This Item:
File | Description | Size | Format | |
---|---|---|---|---|
1995 JESI V44-2 p90.pdf Restricted Access | Restricted Access | 266.6 kB | Adobe PDF | View/Open Request a copy |
Items in RRI Digital Repository are protected by copyright, with all rights reserved, unless otherwise indicated.